Measuring Force of Scanning Probes of Coordinate Machines

Authors: Maciej Byszewski¹² and Adam Woźniak¹
¹ Warsaw University of Technology, Faculty of Mechatronics, Institute of Metrology and Biomedical Engineering; 8 A. Boboli St., 02-525 Warsaw, Poland
² Multisensor sp. z o.o. 80/82 Grzybowska St., 00-844 Warsaw, Poland

This article is a rewritten summary and analysis of the original paper published in Metrology and Measurement Systems, Vol. 32 (2025) No. 3, pp. 1–11, DOI: 10.24425/mms.2025.155806. All credit goes to the original authors, Maciej Byszewski and Adam Woźniak. The original paper can be accessed here. Quotes from the authors are included to preserve their voice and insights.

Abstract

As the authors state: “The paper presents a new method, research setup and the results of the measuring force tests of active and passive scanning probes used in coordinate measuring machines. On the basis of the experimental tests, differences in the characteristics of the pressure of the measuring tip on the measured surface of the active and passive probes are described. These differences concern both the variability of the measuring force and the stabilisation time of the pressure force during the measurement.”

The research highlights the importance of understanding measuring forces in coordinate measurements, particularly for non-rigid workpieces. Keywords: coordinate measurements, scanning probes, and measurement force.

1. Introduction

Coordinate measurement technology is a cornerstone in industries like automotive, aviation, and precision manufacturing. The authors note: “The number of coordinate measuring machines (CMM) operating in measurement laboratories and production halls is constantly growing. This is influenced by the progressive automation of the production of machine parts and devices and with it the need to adjust the rhythm of dimensional control and deviations of shape and position.”

Scanning probes, unlike triggering probes, allow continuous contact with the measured surface, reducing measurement time. The paper distinguishes between active probes (with force generators for quasi-constant force) and passive probes (using springs). The authors emphasize: “The effect of these forces on the measurement error is inversely proportional to the stiffness of the measured element but also depends on the stiffness of the measuring stylus, the diameter of the measuring stylus, and the curvature of the measured workpiece.”

The goal is to present complete measuring force characteristics for active and passive probes, addressing gaps in existing literature focused mainly on triggering probes.

2. Experimental Setup

The setup includes a positioning unit, computer, force sensor amplifier, nanopositioner controller, and data acquisition card. As described by the authors: “The executive unit consists of a measuring surface, which is a reference plate attached to a Kraftsensoren Kd24s force sensor with a measuring range of ±2 N. The measuring surface can be moved in the normal direction by a 3D Physik Instrumente P-611.3S nano-positioner with a measuring range of 120 μm in each axis.”

The setup allows testing in various directions using a tilting base and rotary table.

For illustration, here’s an example of an experimental setup for testing measuring forces in CMM scanning probes:

3. Calibration of the Setup and Determination of Measurement Uncertainty

Calibration used M1 class weights (20g, 50g, 100g, 200g). Linear regression showed R² = 99.9688%.

Uncertainty is calculated as:
u = √(u_c² + u_k² + u_w² + u_o²)

Where:
u_c = 0.002 N (force sensor uncertainty)
u_k = 3.84 × 10^{-7} N (data acquisition card)
u_w = 0.001 N / √3 = 0.00058 N (amplification factor)
u_o = 0.001 N (weights)

Final uncertainty: u = √(0.002² + 3.84 × 10^{-14} + 0.00058² + 0.001²) = 0.0023 N

4. Results

Tests were conducted on a Zeiss Accura CMM with VAST Gold (active) and VAST XXT (passive) probes. Active probe force set to 200 mN; passive depends on deflection.

Measurements varied by stylus length (L), angles α (0°-360°, 30° increments) and β (0°-90°, 22.5° increments), in static and dynamic modes.

Force Characteristics

Examples of time-domain characteristics:

For active probe:

The authors observe three phases: pre-contact (I), contact initiation and stabilization (II), stabilized force (III). “In the case of the active probe tested, Phase II… lasts almost 2 seconds.” For passive: “stabilises much faster, after less than 1 second.”

Parameters: F_max (max force), F_avg (average stabilized), 2s (double std dev), R (range in dynamic).

Angular characteristics are quasi-circular. For active probes, minimal differences with stylus length; for passive, force decreases with longer stylus due to moment arm.

Polar plots illustrate forces at different angles:

For active probe:

For passive probe:

In dynamic mode, force variability is high (±50% for active, up to 3x for passive).

Schematics of probes:

Active probe with flat parallel springs:

Passive probe with diaphragm spring:

5. Conclusions

The authors conclude: “In the case of each contact measurement, the start of the measurement… is of an impact nature. In this phase, in the case of the active probe, the measuring force value momentarily exceeds the set value by almost eight times, and the stabilisation time of the measuring force is almost 2 seconds. In the case of the passive probe, stabilisation… takes less than half the time, however, the value of the measuring force… can differ many times.”

Future work: Impact on measurement error via stylus deflections and Hertzian deformations.

Acknowledgements

The authors thank Dr. Michał Jankowski for measurement assistance.

References

[1] Holub, M., Jankovych, R., Andrs, O., & Kolibal, Z. (2018). Capability assessment of CNC machining centres as measuring devices. Measurement, 118, 52–60. https://doi.org/10.1016/j.measurement.2018.01.007

[2] Gao, W., Haitjema, H., Fang, F., Leach, R., Cheung, C., Savio, E., & Linares, J. (2019). On-machine and in-process surface metrology for precision manufacturing. CIRP Annals, 68(2), 843–866. https://doi.org/10.1016/j.cirp.2019.05.005

[3] Jankowski, M., & Woźniak, A. (2018). Testing of the delay time of wireless communication of CNC machine tools’ probes and controller. Metrology and Measurement Systems, 25(2), 377–386. https://doi.org/10.24425/119559

[4] Achiche, S., & Wozniak, A. (2011). Three-dimensional modeling of coordinate measuring machines probing accuracy and settings using fuzzy knowledge bases: Application to TP6 and TP200 triggering probes. Artificial Intelligence for Engineering Design Analysis and Manufacturing, 26(4), 425–441. https://doi.org/10.1017/s0890060411000151

[5] Sepahi-Boroujeni, S., Mayer, J., & Khameneifar, F. (2021). Efficient uncertainty estimation of indirectly measured geometric errors of five-axis machine tools via Monte-Carlo validated GUM framework. Precision Engineering, 67, 160–171. https://doi.org/10.1016/j.precisioneng.2020.09.027

[6] Gąska, A., Sładek, J., Ostrowska, K., Kupiec, R., Krawczyk, M., Harmatys, W., Gąska, P., Gruza, M., Owczarek, D., Knapik, R., & Kmita, A. (2015). Analysis of changes in coordinate measuring machines accuracy made by different nodes density in geometrical errors correction matrix. Measurement, 68, 155–163. https://doi.org/10.1016/j.measurement.2015.02.056

[7] Corrado, A., & Polini, W. (2019). Measurement of high flexibility components in composite material by touch probe and force sensing resistors. Journal of Manufacturing Processes, 45, 520–531. https://doi.org/10.1016/j.jmapro.2019.07.038

[8] Ascione, R., & Polini, W. (2010). Measurement of nonrigid freeform surfaces by coordinate measuring machine. The International Journal of Advanced Manufacturing Technology, 51, 1055–1067. https://doi.org/10.1007/s00170-010-2684-5

[9] Forrest, E. C., Mertes, R., Gray, J. M., Brumbach, M. T., Ramsdale, S. J., Argibay, N., & Tran, H. D. (2019). Plastic deformation and material transfer on steel gage blocks during low force mechanical probing. Precision Engineering, 57, 253–261. https://doi.org/10.1016/j.precisioneng.2019.04.013

[10] Wozniak, A., Byszewski, M., & Jankowski, M. (2013). Setup for triggering force testing of touch probes for CNC machine tools and CMMs. Measurement Science Review, 13(1), 29–33. https://doi.org/10.2478/msr-2013-0004

[11] Park, J., Kwon, K., & Cho, N. (2006). Development of a coordinate measuring machine (CMM) touch probe using a multi-axis force sensor. Measurement Science and Technology, 17(9), 2380–2386. https://doi.org/10.1088/0957-0233/17/9/002

[12] Li, Y., Zeng, L., Tang, K., & Li, S. (2019). A dynamic pre-travel error prediction model for the kinematic touch trigger probe. Measurement, 146, 689–704. https://doi.org/10.1016/j.measurement.2019.07.005

[13] Woźniak, A., & Męczyńska, K. (2020). Measurement hysteresis of touch-trigger probes for CNC machine tools. Measurement, 156, 107568. https://doi.org/10.1016/j.measurement.2020.107568

[14] Rępalska, M., Woźniak, A., & Łojko, S. (2024). Reliability of probes for CNC machine tools. Precision Engineering, 88, 767–776. https://doi.org/10.1016/j.precisioneng.2024.04.020

[15] Li, Z., Shen, Y., Zhang, W., Huang, N., Zhu, L., & Zhang, Y. (2024). Five-Axis continuous sweep scanning for curvature variable surfaces. IEEE Transactions on Instrumentation and Measurement, 73. https://doi.org/10.1109/tim.2024.3485402

[16] Forrest, E. C., Mertes, R., Gray, J. M., Brumbach, M. T., Ramsdale, S. J., Argibay, N., & Tran, H. D. (2019b). Plastic deformation and material transfer on steel gage blocks during low force mechanical probing. Precision Engineering, 57, 253–261. https://doi.org/10.1016/j.precisioneng.2019.04.013


Previous
Previous

Understanding the GUM Methodology: Applications to Point Clouds in Coordinate Metrology

Next
Next

Hybrid Machine Learning Framework for Predicting Geometric Deviations from 3D Surface Metrology